ATC 6900 GC是美*安普科技中心*新推出的气相色谱仪,该仪器以强大的APC技术(Advanced pneumatic control****控制)**控制整体系统的流量;**的指尖触摸直接操作直观可感知的操作界面;强大的网络(LAN)通讯加速了数据处理速度。
*际通用的强大ATC-Clarity色谱数据处理及仪器控制软件。该软件符合21CFR Part 11.
ATC 6900 GC色谱主机与Tekmar公司的StratumPTCTM吹扫捕集浓缩仪、AquatekTM100液体自动进样器、AtomxTM固/液全自动吹扫捕集仪、VersaTM顶空自动进样器、HT3TM静态/动态顶空自动进样器等前处理设备,可**、轻松地完成您的分析工作。
A. Features(仪器特征)
触摸显示直观图形界面
Intuitive graphic user interface with large touchpad color LCD (5.7”).
强大的APC使分析更*确
Strongly enhanced powerful APC for accurate and precise results
自动存储20种分析方法
Automatic loading and saving up to 20 different analytical methods.
*速的网络数据传输
High speed data process by network (LAN) communication.
可安装多种进样口和检测器
Various inlets and detectors installable up to three per each.
自动关机
Time control : Automatic turn-on/off
10个加热区
Ten heated zone :
1 heated oven zone
3 heated injector zone
3 heated detector zone
2 heated valve zone
1 heated methanizer zone
B. Specifications (规格详述)
1. Gas Chromatograph ATC 6900 GC Oven System Module(主机系统)
柱箱容积:14L
Usable volume : 14L
自动控制冷却控制
Automatic cooling under processor control
温度控制范围:环境温度4-450℃
Temperature operating range : 4℃ above ambient to 450℃ or more
25平台,26阶程序升温
Programming: 25 ramps, 26 steps temperature program
*大升温速率:100℃/min
Maximum set-point heating rate: 100℃/min
冷却速率:450℃ 降*50℃ 需6.5分钟
Cool down rate: 6.5 minutes from 450℃ to 50℃
控温*度:±0.01℃ (等度), ±0.1℃ (梯度)
Temperature stability : ±0.01℃ (Isothermal), ±0.1℃ (Gradient)
2. Injector (Maximum no. of inlet installation: three)(注样器系统,可安装3个)
(1) Packed inlet(填充柱进样系统)
**设定温度:450
Maximum Temperature : 450 oC
**设定范围:
Pressure setting range : 0.01 ~ 100 psi
总流量设定范围:
Total flow setting range : 0.1~100 ml/min
(2) Capillary inlet (Split/Splitless)(**管进样系统(分流/不分流))
**设定温度
Maximum Temperature : 400oC
**设定范围:
Pressure setting range : 0.01 ~ 100 psi
总流量设定范围:
Total flow setting range :
0.1 ~ 400 ml/min N2
0 ~ 1000ml/min He
3. Advanced Pneumatic Control(**控制)
6个APC模块控制注样器、检测器和辅助气
所有气体均有APC控制
All gas flows controlled by APC
程序流量:5阶
Programmed Flow : 5step
程序**:5阶
Programmed Pressure : 5step
检漏测试
Leak detection
检漏报警后,自动*行到关闭状态
After an alarm for a shortage of gas, it automatically saves the method to method no. 0 and shuts down
4.Detectors(检测器)
- Maximum no. of detector installation: three(可安装3个检测器)
- Data Acquisition Rate : 200 Hz (采样频率)
(1) Flame Ionization Detector(FID)
(2) Thermal Conductivity Detector(TCD)
MDL : 2.5 ng/ml (Standard), 400 pg/ml (uTCD)
(3) Nitrogen-Phosphorous Detector(NPD)
(4) Flame Photometric Detector(FPD)
(5) Pulsed Discharge Detector(PDD )
Helium ionization mode(PDHID) :MDL (organic compound : low ppb;permanent gas : low ppm)
Electron capture mode(PDECD) :MDL 10-15
(6) Electron Capture Detector (ECD</str